| Circulation Method | Forced convection |
| Temp. Setting Range | Room temp. +10~210°C |
| Temp. Control Accuracy | ±1°C (at 210°C) |
| Temp. Fluctuation | ±1°C (at 210°C) |
| Temp. Distribution Accuracy | ±3.5°C (at 210°C) |
| Temp. Rising Time | ~60min. |
| Interior / Exterior Material | Stainless Steel / Chrome free electrogalvanized carbon steel sheet coated with chemical-proof baked-on finish |
| Insulation Material | Glass wool |
| Heater | SUS 2.4kW |
| Sensor | K type Thermocouple |
| Fan Type / Fan Motor | Sirocco Fan / Condenser type motor 30W |
| Cable Port | I.D. 33mm (right side) |
| Exhaust Port | I.D. 33mm x 2 (back) |
| Temperature Control | PID control by microprocessor |
| Temperature Display | Temp. display: Green 4-digit LED Digital Display (increment: 1°C)Setting temp. display: Orange 5-digit LED Digital Display (increment: 1°C) |
| Timer | 0 min~99 hrs 59 min (increment: 1 min. or 1 hr.) |
| Heater Control | Triac with Zero-cross control |
| Operation Function | Fixed temperature, Auto start, Auto stop, Quick auto stop, Program (99 steps, 99 patterns, repeat operation function) |
| Additional Function | Power on and operation time accumulation function (up to 65535 hours), Calendar time (24 hours), Calibration offset, Accumulated power consumption monitoring, Total CO ) emission and heater output, Power recovery mode, Save and access operator setting information, key lock |
| Safety Device | Self-diagnostic functions (Sensor failure, SSR short circuit, Heater failure, Main relay contact failure, Automatic overheat prevention), earth leakage breaker, Independent overheat prevention device |
| Internal Dimensions WxDxH | 600 x 500 x 1000 mm |
| External Dimensions WxDxH | 710 x 650 x 1640 mm |
| Internal Capacity | 300L |
| Shelf Plate With Standard Load | ~15kg/pc |
| Shelf Rest Step Number / Pitch | 29 steps / 30mm |
| Power Source | 220V 11.5A (no plug, round terminal) |
| Weight | ~108kg |
| Shelf Plate / Bracket | Stainless steel punched metal |
| 4 pcs. / 8 pcs. |