G100 CO2 INCUBATOR ANALYZER |
| POWER SUPPLY |
| Battery type | Li Ion |
| Battery life | 10 Hours (8 hours with pump) |
| Battery lifetime | >300 Cycles |
| Battery charger | 5v DC external power supply and internal charging circuit |
| Charge time | 3 Hours |
| Alternative power | USB connector DC power supply |
| GAS RANGES |
| Gases Measured | CO2 | By custom dual wavelength infra-red cell with reference channel |
| O2 (Optional) | By internal electrochemical cell |
| Oxygen cell lifetime | Approximately 3 years in air |
| Range | CO2 O2 | 0-20% 0-100% |
Measurement Accuracy | Gas |
CO2 O2 | ± 1% of range after calibration ± 1% of range after calibration |
| Response time, T(90) | CO2 O2 | ≤ 20 seconds ≤ 60 seconds |
| * plus accuracy of calibration gas used |
| FACILITIES |
| Temperature (Optional) | X2 using optional probes, range 32 °F to 122 °F (0 °C to +50 °C) |
| Temperature accuracy, typical | ± 0.2 °C over the range (± 0.1 °C between 32-44 °C) |
| Barometric pressure | 800-1200 mbar |
| RH measurement (Optional) | RH Probe 0-100% RH non condensing |
| RH accuracy | ±1.5%RH across the range |
| Visual and audible alarm | User selectable CO2 and O2 alarm levels |
| Communications | USB type B mini-connector, HID device class |
| Data Storage | 1000 reading sets + 270 events |
| PUMP |
| Flow | 100cc/min typically |
| ENVIRONMENTAL CONDITIONS |
| Operating temperature | 41 °F to 104 °F (5 °C to +40 °C) |
| Relative humidity | 0 - 95% non condensing |
| Barometric pressure | ±201 H2O from calibration pressure, (±500mbar) |
| PHYSICAL |
| Weight | 1.1 lbs 495 grams |
| Size | L 6.5 in, W 3.9 in, D 2.2 in (L 165mm, W100mm, D55mm) |
| Case Material | ABS / Polypropylene with Silicone Rubber Inserts |
| Keys | 17 Resin capped Silicone rubber keys |
| Display | Liquid crystal display, 128 x 64 pixel With RGB LED back-light |
| Gas sample filters | Built-in gas dryer tube to remove moisture User replaceable PTFE water trap filter |