| Operating temperature | -20°C to +60°C |
| Calibration temperature | 20…28°C |
| Relative humidity | < 95% RH (non-cond.) |
| Protection degree (mounted in panel) | IP65 / Type 4X, UL50E |
| Dimensions (HxWxD) | 48 x 96 x 120 mm |
| Cut out dimensions | 44.5 x 91.5 mm |
| Weight approx. | 230 g |
| Wire size, pin 41-46 (max.) | 1 x 1.5 mm2 stranded wire |
| Wire size, others, max. | 1 x 2.5 mm2 stranded wire |
| Vibration | IEC 60068-2-6 |
| 2…13.2 Hz | ±1 mm |
| 13.2…100 Hz | ±0.7 g |
| Supply voltage, universal | 21.6…253 VAC, 50…60 Hz or 19.2…300 VDC |
| Max. required power | 2.5 W (5714A) 3.0 W (5714B/C) 3.5 W (5714D) |
| Internal power dissipation | 2.2 W (5714A) 2.7 W (5714B/C) 3.2 W (5714D) |
| Isolation voltage, test / working | 2.3 kVAC / 250 VAC |
| Temperature input, programmable (0…90%, 100…10%) | 1…60 s |
| mA / V input (programmable) | 0.4…60 s |
| 2-wire supply (pin 46…45) | 25…15 VDC / 0…20 mA |
| Signal / noise ratio | Min. 60 dB (0…100 kHz) |
| Accuracy | Better than 0.1% of selected range |
| EMC immunity influence | < ±0.5% of readout |
| RTD type | Pt10/20/50/100/200/250; Pt300/400/500/1000; Ni50/100/120/1000; Cu10/20/50/100 |
| Cable resistance per wire (max.) | 50 Ω |
| Sensor current | Nom. 0.2 mA |
| Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
| Linear resistance min….max. | 0 Ω…10000 Ω |
| Potentiometer min….max. | 10 Ω…100 kΩ |
| Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5, LR |
| CJC via internally mounted sensor | ±(2.0°C + 0.4°C * Δt) |
| Δt = | Internal temp.-ambient temp. |
| Sensor error detection | Yes |
| Sensor error current: When detecting / else | Nom. 2 μA / 0 μA |
| Measurement range | 0…20 mA |
| Programmable measurement ranges | 0…20 and 4…20 mA |
| Input resistance | Nom. 20 Ω + PTC 25 Ω |
| Sensor error detection | Loop break 4…20 mA |
| Measurement range | 0…12 VDC |
| Programmable measurement ranges | 0/0.2…1; 0/2…10 VDC |
| Input resistance | Nom. 10 MΩ |
| Display readout | -1999…9999 (4 digits) |
| Decimal point | Programmable |
| Digit height | 13.8 mm |
| Display updating. | 2.2 times / s |
| Input outside input range is indicated by | Explanatory text |
| Signal range | 0…20 mA |
| Programmable signal ranges | 0…20/4…20/20…0/20…4 mA |
| Load (at current output) | ≤ 800 Ω |
| Load stability | ≤ 0.01% of span / 100 Ω |
| Sensor error indication | 0 / 3.5 / 23 mA / none |
| NAMUR NE43 Upscale/Downscale | 23 mA / 3.5 mA |
| Output limitation, on 4…20 and 20…4 mA signals | 3.8…20.5 mA |
| Output limitation, on 0…20 and 20…0 mA signals | 0…20.5 mA |
| Current limit | ≤ 28 mA |
| Relay functions | Setpoint |
| Hysteresis | 0…100% |
| ON and OFF delay | 0…3600 s |
| Sensor error reaction | Break / Make / Hold |
| Max. voltage | 250 VRMS |
| Max. current | 2 AAC |
| Max. AC power | 500 VA |
| Max. load at 24 VDC | 1 A |
| EMC | 2014/30/EU |
| LVD | 2014/35/EU |
| EAC | TR-CU 020/2011 |
| DNV-GL Marine | Stand. f. Certific. No. 2.4 |
| EU RO Mutual Recognition Type Approval | MRA000000Z |
| UL | UL 508 |