| Operating temperature | -40°C to +85°C |
| Calibration temperature | 20...28°C |
| Relative humidity | < 95% RH (non-cond.) |
| Protection degree (encl./terminal) | IP68 / IP00 |
| Dimensions | Ø 44 x 20.2 mm |
| Weight approx | 50 g |
| Wire size | 1 x 1.5 mm2 stranded wire |
| Screw terminal torque | 0.4 Nm |
| Vibration | IEC 60068-2-6 |
| 2...25 Hz | ±1.6 mm |
| 25...100 Hz | ±4 g |
| Supply voltage | 7.2...30 VDC |
| Internal power dissipation | 25 mW…0.7 W |
| Isolation voltage, test / working | 1.5 kVAC / 50 VAC |
| Response time (programmable) | 1...60 s |
| Voltage drop | 7.2 VDC |
| Warm-up time | 5 min. |
| Programming | Loop Link |
| Signal / noise ratio | Min. 60 dB |
| EEprom error check | < 3.5 s |
| Accuracy | Better than 0.05% of selected range |
| Signal dynamics, input | 20 bit |
| Signal dynamics, output | 16 bit |
| Effect of supply voltage change | < 0.005% of span / VDC |
| EMC immunity influence | < ±0.5% of span |
| Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% of span |
| Max. offset | 50% of selected max. value |
| RTD type | Pt100, Ni100, lin. R |
| Cable resistance per wire | 5 Ω (max.) |
| Sensor current | Nom. 0.2 mA |
| Effect of sensor cable resistance (3-/4-wire) | < 0.002 Ω / Ω |
| Sensor error detection | Yes |
| Linear resistance min....max | 0 Ω...5000 Ω |
| Thermocouple type | B, E, J, K, L, N, R, S, T, U, W3, W5, LR |
| Cold junction compensation (CJC) | < ±1.0°C |
| Sensor error detection | Yes |
| Sensor error current: When detecting / else | Nom. 33 μA / 0 μA |
| Measurement range | -12...800 mV |
| Min. measurement range (span) | 5 mV |
| Input resistance | 10 MΩ |
| Signal range | 4…20 mA |
| Min. signal range | 16 mA |
| Load (at current output) | ≤ (Vsupply - 7.2) / 0.023 [Ω] |
| Load stability | ≤ 0.01% of span / 100 Ω |
| Sensor error indication | Programmable 3.5…23 mA |
| NAMUR NE43 Upscale/Downscale | 23 mA / 3.5 mA |
| Updating time | 440 ms |
| of span | = of the presently selected range |
| ATEX | II 1 G Ex ia IIC T4...T6 Ga, II 1 D Ex ia IIIC Da, I M1 Ex ia Ma |
| IECEx | Ex ia IIC T4...T6 Ga, Ex ia IIIC Da, Ex ia I Ma |
| FMus | Cl. I, Div. 1, Gp. A, B, C, D T4/T6; Cl. I Zone 0, AEx ia IIC T4/T6; Cl. 1, Div. 2, Gp. A, B, C, D, T4/T6 |
| CSA | Cl. I, Div. 1, Gp. A, B, C, D Ex ia IIC, Ga |
| INMETRO | Ex ia IIC T6...T4 Ga, Ex ia IIIC Da |
| EMC | 2014/30/EU |
| RoHS | 2011/65/EU |
| EAC | TR-CU 020/2011 |
| DNV-GL Marine | Stand. f. Certific. No. 2.4 |
| ATEX 2014/34/EU | KEMA 06ATEX0062X |
| IECEx | DEK 13.0035X |
| INMETRO | DEKRA 13.0001 X |
| CCOE | P337392/2 |