| Environmental Conditions | |
| Operating temperature | -25°C to +70°C |
| Storage temperature | -40°C to +85°C |
| Calibration temperature | 20...28°C |
| Relative humidity | < 95% RH (non-cond.) |
| Protection degree | IP20 |
| Installation in | Pollution degree 2 & meas. / overvoltage cat. II |
| Mechanical specifications | |
| Dimensions (HxWxD) | 113 x 6.1 x 115 mm |
| Weight approx. | 70 g |
| DIN rail type | DIN EN 60715/35 mm |
| Wire size | 0.13...2.5 mm2 / AWG 26...12 stranded wire |
| Screw terminal torque | 0.5 Nm |
| Vibration | IEC 60068-2-6 |
| 2...25 Hz | ±1.6 mm |
| 25...100 Hz | ±4 g |
| Supply voltage | 16.8...31.2 VDC |
| Max. required power | 0.52 W |
| Max. power dissipation | 0.52 W |
| Response time | |
| Response time (0...90%, 100...10%) | < 30 ms / 300 ms (selectable) |
| Signal / noise ratio | Min. 60 dB |
| Programming | DIP-switches |
| Signal dynamics, input | 23 bit |
| Signal dynamics, output | 18 bit |
| Accuracy | Better than 0.1% of sel. range |
| EMC immunity influence | < ±0.5% of span |
| Extended EMC immunity: NAMUR NE21, A criterion, burst | < ±1% of span |
| Incorrect DIP-switch setting identification | 0 V / 0 mA output; LED 0.5 s / 1 Hz |
| Input specifications | < ±0.5% of span |
| TC input | < ±1% of span |
| Temperature range, TC J | -100...+1200°C |
| Temperature range, TC K | -180...+1372°C |
| Min. measurement range (span) - TC J & K | 50°C |
| Accuracy: the greater of | Better than 0.1% of span or 1°C |
| Temperature coefficient: the greater of | 0.1°C/°C or ≤ ±0.01%/°C |
| Sensor cable resistance | < 5 kΩ per wire |
| Cold junction compensation (CJC): Accuracy at internal CJC | Better than ±2.5°C |
| Internal CJC error detection | Yes |
| Open Thermocouple detection | Yes - selectable via DIP-switch |
| Output specifications | < 0.002 Ω / Ω |
| Common output specifications | Yes - selectable via DIP-switch |
| Updating time | 10 ms |
| Current output | < 18 Ω |
| Signal range | 0...23 mA |
| Programmable signal ranges | 0 / 4...20 mA |
| Sensor error indication (0...20 mA) | 0 mA or 23 mA / OFF |
| Sensor error indication (4...20 mA) | 3.5 mA or 23 mA / acc. to NAMUR NE43 or OFF |
| Load ( at current output) | ≤ 600 Ω |
| Load stability | ≤ 0.01% of span / 100 Ω |
| Current limitation at low output load | < 60 mA peak / < 4 mA average |
| Voltage output | 3.5 mA or 23 mA / acc. to NAMUR NE43 or OFF |
| Programmable signal ranges | 0/1...5 and 0/2...10 V |
| Sensor error indication | 0 V / 10% above the max. / none |
| Load ( at voltage output) | ≥ 10 kΩ |
| Open output | < 18 V |
| I.S. / Ex marking | 0/1...5 and 0/2...10 V |
| ATEX | II 3 G Ex ec IIC T4 Gc |
| IECEx | Ex ec IIC T4 Gc |
| FM, US | Cl. I, Div. 2, Gp. A, B, C, D T4 or Cl. I, Zone 2, AEx nA IIC T4 |
| FM, CA | Cl. I, Div. 2, Gp. A, B, C, D T4 or Cl. I, Zone 2, Ex nA IIC T4 |
| EAC Ex | 2Ex nA IIC T4 Gc X |
| Observed authority requirements | Cl. I, Div. 2, Gp. A, B, C, D T4 or Cl. I, Zone 2, AEx nA IIC T4 |
| EMC | 2014/30/EU |
| LVD | 2014/35/EU |
| ATEX | 2014/34/EU |
| RoHS | 2011/65/EU |
| EAC | TR-CU 020/2011 |
| EAC Ex | TR-CU 012/2011 |
| Approvals | 2011/65/EU |
| ATEX | KEMA 10ATEX0147 X |
| IECEx | KEM 10.0068X |
| c FM us | FM17US0004X / FM17CA0003X |
| c UL us, UL 61010-1 | E314307 |
| EAC Ex | RU C-DK.HA65.B.00355/19 |
| DNV Marine | TAA00001RW |