Display | LCD (5 inches diagonal) |
Backlight | Blue (adjustable duration) |
Over Range Display | OL for pressure, for temperature |
Auto Power Off | 30 minutes of inactivity (adjustable) |
Maximum Manifold Pressure | 800 Psig (5500 kPa) |
Battery Type | 6 x AA alkaline |
Battery Life | 350 hours typical (without vacuum, backlight, and wireless) |
Radio Frequency | 2.4 GHz |
Wireless Range | 1000 feet (305 meters) line of sight (Distance decreases through obstructions) |
Data Port | Micro USB (for extracting logged data or updating firmware) |
Operating Environment | 32°F to 122°F (0°C to 50°C) at <75% RH |
Storage Environment | -4°F to 140°F (-20°C to 60°C) at <80% RH (with battery removed) |
Temperature Coefficient | 0.1 x (specified accuracy) per °C (0°C to 18°C, 28°C to 50°C), per 1.8°F (32°F to 64°F, 82°F to 122°F) |
Weight | 4.03 lbs (1.83 kg) |
Water Resistant | Designed to IP54 |
Pressure |
Sensor Type | Absolute pressure sensors |
Port Type | (3) 1/4" standard NPT male flare fittings |
Pressure Range and Units | 580 Psig (English), 40.00 Bar (Metric), 4.000 MPa (Metric), and 4000 kPa (Metric) |
Negative Pressure Range and Units | 29 inHg (English), 74 cmHg (Metric) |
Resolution | 0.1 Psig; 0.01 Bar; 0.001 MPa; 1 kPa; 0.1 inHg; 1 cmHg |
Negative Pressure Accuracy | 29 inHg to 0 inHg: ±0.2 inHg; 74 cmHg to 0 cmHg: ±1 cmHg |
Pressure Accuracy | - 0 Psig to 200 Psig: ±1 Psig;
- 200 Psig to 580 Psig: ±(0.3% of reading + 1 Psig);
- 0 Bar to 13.78 Bar ±0.07 Bar;
- 13.78 Bar to 40.00 Bar: ±(0.3% of reading + 0.07 Bar);
- 0 MPa to 1.378 MPa: ±0.007 MPa;
- 1.378 MPa to 4.000 MPa: ±(0.3% of reading + 0.007 MPa);
- 0 kPa to 1378 kPa: ±7 kPa;
- 1378 kPa to 4000 kPa: ±(0.3% of reading + 7 kPa)
|
Temperature |
Sensor Type | Type K thermocouple (nickel chromium/nickel aluminum) |
Jack Type | (3) Type K thermocouple |
Range | -50°F to 257°F (-46°C to 125°C), limited by the thermocouple specification. Display range is -95°F to 999.9°F (-70°C to 537.0°C) |
Resolution | 0.1°F (0.1°C) |
Accuracy | Shown accuracies are after field calibration.- ±(1.0°F) -95°F to 200°F, ±(2.0°F) 200°F to 999.9°F;
- ±(0.5°C) -70°C to 93°C, ±(1.0°C) 93°C to 537.0°C
|
Deep Vacuum |
Sensor Type | Thermistor |
Port Type | (3) 1/4" standard NPT male flare fittings |
Range and Units | - 50 to 9999 microns of mercury (English)
- 6 to 1330 Pascals (Metric)
- 0.06 to 13.3 mBar (Metric)
- 50 to 9999 mTorr (Metric)
- 0.05 to 10 Torr (Metric, equivalent to mmHg)
|
Best Resolution | - 1 micron of mercury (below 2000 microns)
- 1 Pascal (below 250 Pascals)
- 0.001 mBar (below 2.5 mBar)
- 1 mTorr (below 2000 mTorr)
- 0.001 Torr (below 2.5 Torr)
|
Accuracy 77°F (25°C) | - ±(5% of reading + 5 microns of mercury)
- ±(5% of reading + 1 Pascal)
- ±(5% of reading + 0.01 mBar)
- ±(5% of reading + 5 mTorr)
- ±(5% of reading + 0.005 Torr)
|