Subject to the use of all the optional accessories |
UHH2 Specifications |
Display | 4.3" QHD Gorilla glass, 960x540 |
Wireless Distance | 50" (15 m) |
Response Time | 1 second |
Temperature Limits | Operating: -4 to 140°F (-20 to 60°C); Storage: -40 to 176°F (-40 to 80°C) |
Humidity Limits | 5 to 95% RH |
Power Requirements | 2000 mAh lithium ion battery, installed functional, non-replaceable |
Memory | RAM 1G and ROM 4G |
Operating System | Android 4.2.2 |
CPU | MTK6589W Quad Core 1.2 GHz |
Enclosure Rating | IP68 |
Weight | 15.9 oz (450 g) |
Agency Approvals | CE |
AP2 Thermo-Anemometer Probe Specifications |
Service | Dry, clean air |
Temperature Limits | Process: -20 to 200°F (-29 to 93°C); Ambient: -20 to 140°F (-29 to 60°C); Battery Charging: 32 to 113°F (0 to 45°C) |
Range | Air Velocity: 0 to 6000 FPM (0 to 30 m/s); Volumetric Air: 999,999 in selected flow units; Temperature: -40 to 212°F (-40 to 100°C) |
Accuracy | Air Velocity: ±3% FS within temperature range of 40 to 90°F (4 to 32°C); Temperature: ±0.5°F (±0.28°C) |
Response Time | 1 second |
Probe Length | 8" (203 mm) insertion |
Power Requirements | 3.7 V YT562447 Lithium ion battery, installed functional, user replaceable |
Maximum Wireless Distance | 50' (15 m) |
Handle Enclosure | Thermoplastic elastomer over polycarbonate |
Supplied With | Wrist strap |
Weight | 11.2 oz (331.22 g) |
Agency Approvals | CE (not while charging), FCC compliant |
RP2 Thermo-Hygrometer Probe Specifications |
Service | Clean air |
Temperature Limits | Process: -4 to 140°F (-20 to 60°C); Ambient: 5 to 125°F (-15 to 51°C); Battery Charging: 32 to 113°F (0 to 45°C) |
Range | RH: 0 to 100% (non-condensing); Temperature: -22 to 140°F (-30 to 60°C) |
Accuracy | RH: ±2% at 25°C (10 to 90% RH); ±4% (0 to 10, 90 to 100% RH); Temperature: ±0.54°F at 77°F (±0.3°C at 25°C) |
Response Time | 1.5 second |
Probe Length | 8" (203 mm) insertion |
Power Requirements | 3.7 V YT562447 Lithium ion battery, installed functional, user replaceable |
Maximum Wireless Distance | 50' (15 m) |
Handle Enclosure | Thermoplastic elastomer over polycarbonate |
Supplied With | Wrist strap |
Weight | 11.2 oz (331.22 g) |
Agency Approvals | CE (not while charging), RoHS, FCC compliant |
VP2 Vane Thermo-Anemometer Probe Specifications |
Service | Clean air |
Temperature Limits | Process: -4 to 140°F (-20 to 60°C); Ambient: 5 to 125°F (-15 to 51°C) |
Range | Air Velocity: 50 to 5000 fpm (0.25 to 25 m/s); Temperature: -22 to 140°F (-30 to 60°C); Relative Humidity: 0 to 100% RH; Air Volume: 999,999 in selected units |
Accuracy | Air Velocity: 0.25 to 10 m/s: ±1.5% of reading ±20 fpm (±0.1 m/s); 10 to 20 m/s: 1.5% of reading ±40 fpm (±0.2 m/s); 20 to 25 m/s: ±1.5% of reading ±60 fpm (±0.3 m/s); Temperature: ±0.54°F at 77°F (±0.3°C at 25°C); Relative Humidity: ±2% at 77°F (25°C) (10 to 90% RH); ±4% (0 to 10% and 90 to 100%) |
Response Time | Air Velocity: 1 s; Temperature: 1.5 s; Relative Humidity: 1.5 s; Air Volume: 1 s |
Probe Length | 8" (203 mm) insertion |
Power Requirements | 3.7 V YT562447 Lithium ion battery, installed functional, user replaceable |
Battery Charging Limits | 32 to 113°F (0 to 45°C) |
Maximum Wireless Distance | 50' (15 m) |
Vane Material | Anondized aluminum |
Handle Enclosure | Thermoplastic elastomer over polycarbonate |
Supplied With | Wrist strap |
Weight | 13.6 oz (402.20 g) |
Agency Approvals | CE (not while charging), RoHS, FCC compliant |
WDPM Specifications |
Service | Non-corrosive dry gases |
Accuracy | ±0.5% FS span at 25°C (includes non linearity, hysteresis, and non repeatability) |
Pressure Limits | ±10" w.c. (± 2.5 kPa) |
Engineering Units | Pressure: in w.c., ft w.c., in Hg, psi, OzSl, mm w.c., cm w.c., mBar, Pa, hPa, and kPa; Velocity: fpm, mph, kn, m/s, m/h, k/h, and fps; Flow: cfm, gpm, gph, gpd, m3/s, m3/h, lps, lpm, and lph |
Temperature Limits | Compensated: 32 to 140°F (0 to 60°C); Process/Ambient: 14 to 140°F (-10 to 60°C) |
Thermal Effects | ±0.01% FS/°F (±0.02% FS/°C) |
Battery Charging Limits | 32 to 113°F (0 to 45°C) |
Power Requirements | 3.7 V YT562447 lithium ion battery, installed functionally, user replaceable |
Wireless Distance | At least 50' (15 m) |
Connections | Two barbed connections for use with 1/8" (3.18 mm) or 3/16" (4.76 mm) ID tubing |
Weight | 2.5 oz (70.87 g) |
Agency Approvals | CE with CE approved charger, RoHS, FCC |