Method | Decompression-chamber wall heating |
Operating temp, range | 40-200°C |
Operating vacuum range | 101-0.1kPa (760-1Torr) |
Temp, adjustment accuracy | ± 1.0°C (at 200°C) |
Interior | Stainless steel plate |
Exterior | Cold rolled steel plate with chemical proofing coating |
Insulating material | Glassfiber |
Heating method | Decompressed chamber wall direct heating |
Heater power | 6.5kW |
Observation window | Toughened glass + resin protection panel |
Vacuum gauge | Pointer type, -100~0kPa |
Vacuum pump installation room | Yes |
Temp, control | 3 segments PID |
Temp, setting | Use specialized function menu key and up/down key to set |
Temp, display | Measured temp, display: green 4-digit LED digital display |
Setting temp, display: red 4-digit LED digital display |
Timer | 1min-99 hr 59 min and 100 hr-999 hr 50 min (attached with timing wait function) |
Operation function | Fixed temp, auto start, auto stop, program operation |
Program mode | Program operation 3 segments 30 steps (30 stepsx1, 15 stepsx2,10 stepsx3) |
Additional function | Deviation correction, key lock, power outage compensation |
Heater circuit control | SSR driving |
Sensor | K thermocouple (temp, controller and overheat protector) |
Safety device | Self-diagnosis circuit (abnormal temp, sensing, heater disconnection, auto overheat prevention, SSR short circuit), overheat protector, EBL to prevent overcurrent, key lock. etc. |
Internal dimensions (WxDxHmm) | 800x800x800 |
External dimensions (WxDxHmm] | 1020x1020x1850 |
Internal capacity | 512L |
Air exhaust port | NW40 flange |
Air suction port | Rc 3/8 |
Power source (50/60Hz) rated current | AC220V |
Weight | ~450kg |
Accessories | Stainless steel punching plates, 2 pcs. |