| Circulation Method | Forced convection |
| Operating Temp. Range | Room temperature +30 to 260°C |
| Temp. Control Accuracy | ±0.3°C at 260°C |
| Temp. Fluctuation | ±0.5°C at 260°C |
| Temp. Distribution Accuracy | ±2.5°C at 260°C |
| Temp. Gradient | ±10.0°C at 260°C |
| Max. Temp. Reaching Time | ~70 mins. |
| Clean Level | Class100 (when temperature is stable) |
| Interior/Exterior Material | Stainless steel / Cold rolled steel plate with melamine resin baking finish |
| Heat Insulating Material | Glass wool |
| Door | Single side left swing |
| Heater | Stainless steel pipe heater |
| Fan Type | Sirocco fan, Condenser motor 400W |
| Differential Pressure Meter | Analog type (0 ~ 300Pa) |
| Cable Port/ Exhaust Port | Inner diameter: 33mmx1 (right side) / Outer diameter 61mm |
| Filter | Heat resistant HEPA filter (dust-collection efficiency >99.97% up to 0.3μm particle filtering) |
| Caster Wheels / Adjuster | Free swivel caster wheels without stopper / level adjuster (2 at front) |
| Temp. Control/Setting Method | PID V control / Digital setting with keys |
| Temp. Display System | Top screen: green 4-digit digital LED (resolution 1°C), Bottom screen: orange 5-digit digital LED (resolution 1°C) |
| Other Indications | LED indicates temperature patterns for heating/stabilizing/cooling |
| Operation Functions | Constant temperature operation, Programmed operation (Maximum 99 steps, up to 99 patterns, repeat operation function),; Duration/time select operation function (auto start/auto stop/quick auto stop, program operation) |
| Additional Functions | Variable fan speed, Accumulated on time, operation time function (up to 65,535 hours); calibration offset; accumulated power; consumption monitoring, total CO2 emission monitoring, heater output monitoring; power recovery; setting data save and restore |
| Sensor | K type Thermocouple dual sensor (temperature control and independent overheat prevention device sensors) |
| Heater Control | Triac with zero-cross control |
| Safety Device | Self-diagnostic functions (Detection for Temp. Sensor Failure, Triac Short Circuit, Automatic overheating prevention, Fan motor; failure detection, Heater Line Disconnect, Main Relay Contact Damage), Earth Leakage Breaker, Key Lock Function, Independent; overheating prevention device, Phase reversal relay, Door switch |
| Earth Leakage Breaker | 15A |
| Current leak /short circuit/surge protection, rated sensitivity | 30mA |
| Door Switch | Door open: fan motor and heater circuit off, Door close: fan motor and heater circuit on |
| Internal Dimensions WxDxH | 450x450x450 mm |
| External Dimensions WxDxH | 700x1025x1570 mm |
| Internal Capacity | 91L / 3.21 cu. ft. |
| Weight | ~200 kg / ~441 lbs. |
| Shelf Rest / Pitch | 12 steps / 30mm |
| Withstand Load Of Shelf | ~30 kg / shelf |
| Power Supply 50/60Hz (V±10%) | 220V 3 phase 7A |
| Shelf Plate / Bracket | 2 pcs. / 4 pcs. |