Circulation Method | Forced convection |
Operating Temp. Range | Room temperature +30 to 360°C |
Temp. Control Accuracy | ±0.3°C at 360°C |
Temp. Fluctuation | ±0.5°C at 360°C |
Temp. Distribution Accuracy | ±4.0°C at 360°C |
Temp. Gradient | ±20.0°C at 360°C |
Max. Temp. Reaching Time | ~80 mins. |
Clean Level | Class100 (when temperature is stable) |
Interior/Exterior Material | Stainless steel / Cold rolled steel plate with melamine resin baking finish |
Heat Insulating Material | Glass wool |
Door | Single side left swing |
Heater | Stainless steel pipe heater |
Fan Type | Sirocco fan, Condenser motor 400W |
Differential Pressure Meter | Analog type (0 ~ 300Pa) |
Cable Port/ Exhaust Port | Inner diameter: 33mmx1 (right side) / Outer diameter 61mm |
Filter | Heat resistant HEPA filter (dust-collection efficiency >99.97% up to 0.3µm particle filtering) |
Caster Wheels / Adjuster | Free swivel caster wheels without stopper / level adjuster (2 at front) |
Temp. Control/Setting Method | PID V control / Digital setting with keys |
Temp. Display System | Top screen: green 4-digit digital LED (resolution 1°C), Bottom screen: orange 5-digit digital LED (resolution 1°C) |
Other Indications | LED indicates temperature patterns for heating/stabilizing/cooling |
Operation Functions | Constant temperature operation, Programmed operation (Maximum 99 steps, up to 99 patterns, repeat operation function), Duration/time select operation function (auto start/auto stop/quick auto stop, program operation) |
Additional Functions | Variable fan speed, Accumulated on time, operation time function (up to 65,535 hours); calibration offset; accumulated power consumption monitoring, total CO2 emission monitoring, heater output monitoring; power recovery; setting data save and restore |
Sensor | K type Thermocouple dual sensor (temperature control and independent overheat prevention device sensors) |
Heater Control | Triac with zero-cross control |
Safety Device | Self-diagnostic functions (Detection for Temp. Sensor Failure, Triac Short Circuit, Automatic overheating prevention, Fan motor failure detection, Heater Line Disconnect, Main Relay Contact Damage), Earth leakage breaker, Key Lock Function, Independent overheating prevention device, Phase reversal relay, Door switch |
Earth Leakage Breaker | - 20A
- Current leak /short circuit/surge protection, rated sensitivity 30mA
|
Door Switch | Door open: fan motor and heater circuit off, Door close: fan motor and heater circuit on |
Internal Dimensions WxDxH | 600x600x600 mm / 23.62" x 23.62" x 23.62" |
External Dimensions WxDxH | 850x1175x1720 mm / 33.46" x 46.25" x 67.71" |
Internal Capacity | 216L / 7.62 cu.ft. |
Weight | ~270 kg / ~596 lbs. |
Shelf Rest / Pitch | 17 steps / 30mm / 1.18" |
Power Supply 50/60Hz (V±10%) | 220V 3 phase 14A |
Shelf Plate / Bracket | 3 pcs. / 6 pcs. |