Flue gas O2 Technical data |
Measuring range | 0 to 21 Vol.% |
Accuracy | ±0.2 Vol.% |
Resolution | 0.1 Vol.% |
Reaction time t90 | < 20 sec |
Flue gas CO Technical data |
Measuring range | 0 to 8000 ppm |
Accuracy | ±10 ppm or ±10 % of remaining measurement value (0 to 200 ppm) ±20 ppm or ±5 % of remaining measurement value (201 to 2000 ppm) ±10 % of remaining measurement value (2001 to 8000 ppm) |
Resolution | 1 ppm |
Reaction time t90 | < 40 sec |
Flue gas COlow Technical data |
Measuring range | 0 to 500 ppm |
Accuracy | ±2 ppm (0 to 39.9 ppm) ±5 % of remaining measurement value (40 to 500 ppm) |
Resolution | 0.1 ppm |
Reaction time t90 | < 40 sec |
Flue Gas Draft Technical data |
Measuring range | -9.99 to +40 hPa |
Accuracy | ±0.02 hPa or ±5 % of remaining measurement value (-0.50 to +0.60 hPa) ±0.03 hPa (+0.61 to +3.00 hPa) ±1.5 % of remaining measurement value (+3.01 to +40.00 hPa) |
Resolution | 0.01 hPa with fine draught option 0.001 hPa |
Temperature Technical data |
Measuring range | -40° to 2192 °F / -40 to +1200 °C |
Accuracy | ±0.9 °F (32° to 212.0 °F) / ±0.5 °C (0 to +100.0 °C) ±0.5 % of remaining measurement value (Remaining Range) |
Resolution | 0.1 °F (-40° to 1831.8 °F) / 0.1 °C (-40 to +999.9 °C) 1.0 °F (> +1832 °F) / 1 °C (> +1000 °C) |
Efficiency (calculated) Technical data |
Measuring range | 0 to 120 % |
Resolution | 0.1 % |
Flue Gas loss (calculated) Technical data |
Measuring range | 0 to 99.9 % |
Resolution | 0.1 % |
Flue Gas CO2 calculation (calculated from O2) Technical data |
Measuring range | 0 to CO2 max (Display range) |
Accuracy | ±0.2 Vol.% |
Resolution | 0.1 Vol.% |
Reaction time t90 | < 40 sec |
Velocity / Volume flow Technical data |
Measuring range | 0.15 to 3 m/s |
Resolution | 0.1 m/s |
Pressure measurement Technical data |
Measuring range | 0 to +300 hPa |
Accuracy | ±0.5 hPa (0.0 to 50.0 hPa) ±1 % of remaining measurement value (50.1 to 100.0 hPa) ±1.5 % of remaining measurement value (Remaining Range) |
Resolution | 0.1 hPa with fine pressure option 0.01 hPa |
Ambient CO Technical data |
Measuring range | 0 to 500 ppm |
Accuracy | ±5 ppm (0 to 100 ppm) ±5 % of remaining measurement value (> 100 ppm) |
Resolution | 1 ppm |
Ambient CO2 Technical data |
Measuring range | 0 to 1 Vol.% 0 to 10000 ppm |
Accuracy | ±50 ppm or ±2 % of remaining measurement value (0 to 5000 ppm) ±100 ppm or ±3 % of remaining measurement value (5001 to 10000 ppm) |
Gas leak measurement for combustible gases (via gas leak detection probe) Technical data |
Measuring range | 0 to 10000 ppm CH4 / C3H8; Display range |
Accuracy | Signal optical display (LED) audible signal via buzzer |
Reaction time t90 | < 2 sec |
Temperature (via the fine pressure probe) Technical data |
Measuring range | -40° to 2192 °F / -40 to +1200 °C max. (dependent on probe) |
Accuracy | ±0.9 °F (-40° to 212 °F) / ±0.5 °C (-40 to 100 °C) ±0.5 % of remaining measurement value (Remaining Range) plus probe accuracy |
Resolution | 0.1 °F / 0.1 °C |