Measurement parameters (Norm) volume flow |
Selectable units | m3/h; l/min; m3 |
Measuring range (1:300)1 | 0.25 to 75 m3/h |
Accuracy(norm volume flow) | For compressed air quality classes (ISO 8573: particles-humidity-oil) 1-4-1: ±3 % of meas. value ±0.3 % of final value For compressed air quality classes (ISO 8573: particles-humidity-oil) 3-4-4: ±6 % of meas. value ±0,6 % of final value |
Sensor
| Thermal, glass-coated ceramic sensor (calorimetric measurement procedure) |
Response time | <0.1 sec (for damping parameter = 0), delayable via operating menu (0 to 1 sec) |
Temperature Unit | °C |
Measuring range of Temperature | 0 to +60 °C / 32 °F to +140 °F |
Inputs and outputs, Analog outputs |
Output type | 4 to 20 mA (4-wire) freely scalable between zero and measuring range end |
Load | max. 500 Ω |
Pulse output | Consumption quantity counter (value remains available after reset or power cut due to non-volatile memory), value 1 or 10m3, pulse length 0.02 s to 2 s, 24 VDC level |
Switch output | 2 switch outputs, parameterizable (consumption or volume flow-dependent, opener, closer, hysteresis, window), loadablewith max. 20 to 30 VDC or 250 mA each, switch status is displayed via 2 LEDs |
Voltage supply | 19 to 30 V DC |
Current consumption | <100 mA |
Connection | M12 x 1 plug, loadable up to 250 mA, short-circuit-proof (synchronized), reverse-polarity-proof, overload-proof |
General technical data Design |
Material housing | PBT (GF 20%), diecast zinc, silica-free |
Long measurement stretch | 11.81 in |
Pipe diameter | DN 15 (for 1/2" pipes) |
Weight | 1.98 lb |
Display |
Material | 4-figure alphanumerical display, two operating buttons, operating menu, LED (4 x green for phys. units, 3 x yellow fordisplay x 1,000 or switch status) |
Max. display valuenorm volume flow | 90 m3/h |
Temperature display | 0 to +60 °C, measurement error ±2 K, (+32 to +140 °F) |
Parameterization | 2 operating buttons |
Measurement stretch:thread (both sides)/material | R 1/2, outer threadStainless steel 1.4301 |
Protection class | IP 65/III |
EMC | according to guideline 89/336 EEC |
Media contact | Materials stainless or galvanized steel, PEEK, polyester, Viton, anodized aluminium ceramic |
Operating conditions |
Humidity (sensor) | re. humidity <90 %RH |
Operating temperature (housing) | 0 to +60 °C (+32 to +140 °F) |
Storage temperature | -25 to +85 °C (-13 to +185 °F) |
Measurement medium | Compressed air |
Process pressure | PN 16 (max 16bar/232psi) |
Air quality | ISO 8573: recommended classes 1-4-1 |